important regulatory document 11 October 2025


image

Correct risk classification in accordance with Annex VIII of the Medical Devices Regulation Medical Device Regulation (EU) 2017/745 (MDR): Motorized treadmills

EUDAMED EMDN Code: Z129006 Treadmills for physiotherapy and/or diagnostic use

EUDAMED EMDN Code: Z129006 TREADMILLS FOR PHYSIOTHERAPY AND/OR DIAGNOSTIC USES


To all relevant economic operators, manufacturers, importers, EU authorised representatives, distributors, dealers, OEM/PLM, healthcare facilities, medical technicians, customers, notified bodies and authorities, as well as stakeholders


Motorised treadmills have been an indispensable part of healthcare facilities for many decades, with applications in diagnostics and therapy. However, risks and accidents involving serious injuries and rare fatalities are also known to occur with these powerful machines. Nevertheless, the medical benefits clearly outweigh the risks!

This document is dedicated to the correct risk classification by manufacturers to enable all economic operators, authorities and stakeholders involved to assess, plan and implement the responsibilities, necessary actions and risks accordingly.


Motorised treadmills for physiotherapy and/or diagnostic applications must be classified in risk class IIa or IIb in accordance with Rule 9, and a Notified Body must be involved in certification and monitoring. The CE mark requires a 4-digit identification number from the Notified Body.


Risk class

Rule

CE mark

Notified body

Assessment

Assessment

IIa or IIb

9

image

or other number

yes

correct


300+ kostenlose Häkchen-Symbole & Haken-Bilder - Pixabay

I

13

Wo gehört das CE-Zeichen hin? Klarheit ...

No

incorrect

Rotes Kreuz-Häkchen. Ablehnung und kein Abstimmungssymbol | Premium Vektor


The MDR stipulates that incorrect classification of a medical device means that the product is not marketable and therefore may not be distributed or placed on the European market.


The applicability of Rule 9, i.e. the delivery or exchange of energy when using treadmills as medical devices, is not only based on the intended purpose, the logic and the physical and biomechanical principles of walking and running on a treadmill, but has been confirmed countless times not only by the treadmill manufacturer h/p/cosmos but also by independent authorities and/or bodies and experts from industry and science, as follows:


28 February

2025

BfArM decision dated 28 February 2025 with reference number: 91.3.32-5634-K-0049/24

In a letter dated 11 October 2024, h/p/cosmos, as the manufacturer, submitted an application to the BfArM (Federal Institute for Drugs and Medical Devices in Germany) to decide on the classification of a motorised treadmill with an interface on the basis of Section 6 (2) MPDG (German Medical Device Law Implementation Act).


At the request of h/p/cosmos, the BfArM confirmed that, when used as intended, the treadmill exchanges energy with the patient.

  1. Classification rule 9 is applicable

  2. Mechanical kinetic energy is transferred/exchanged on the treadmill

  3. The treadmill in question is classified as risk class IIa […]


image

image

h/p/cosmos sports & medical gmbh Am Sportplatz 8 DE 83365 Nussdorf-Traunstein / Germany EUDAMED SRN: DE-MF-000006147 phone +49 86 69 86 42 0 fax +49 86 69 86 42 49 email@hpcosmos.com www.hpcosmos.com


Abbreviated excerpt from the BfArM decision of 28 February 2025 for a "simple and completely normal" treadmill without a Robowalk gait trainer, without perturbation and without high speeds, with the following specifications: motor drive, running surface 150/50 cm, speed 0 ... 22 km/h, incline angle 0 ... 25%, interface for ECG and/or spiroergometry, optional heart rate measurement via accessories (e.g. POLAR),

optional adjustable handrails, treadmill without measuring function, treadmill without diagnostic function, application for physiotherapy and diagnostics.

1998 to 2025

Every year

Notified body: TÜV SÜD Product Service GmbH, Munich

Certification audit and annual surveillance audit from 1998 according to MDD / MPG and since 2022 according to MDR / MPDG / ISO 13485, including review of technical documentation with confirmation of applicability of Rule 9. https://www.hpcosmos.com/en/contact-support/media-downloads/certificates

13 September

2024

Bavarian State Office for Health and Food Safety, Munich

Free sale certificate No. AP-2697-22-V6032-46309/12024 in accordance with Art. 60 Regulation (EU) 2017/745 in conjunction with § 10 Medical Devices Implementation Act (MPDG) in the currently applicable versions for submission to the competent authorities.

... after respective review of the EU declarations of conformity https://www.hpcosmos.com/en/contact-support/media-downloads/certificates

Also confirmed by free sales certificates, e.g. with dates including: 8 May 2002, 28 June 2007, 2 May 2013,

18 July 2014, 1 September 2022, 29 August 2023

18 December

2023

Paris Lodron University of Salzburg / Univ.-Prof. Dr. Hermann Schwameder

Statement on energy exchange and expenditure when walking and running on a treadmill

Paris Lodron University of Salzburg, Sports and Movement Science, Head of Department, Head of Biomechanics Working Group, Schlossallee 49, 5400 Hallein-Rif, Austria download

30

University Hospital Halle (Saale) / Associate Professor Dr. phil. Rene Schwesig

Assessment of energy exchange during exercise on treadmill ergometers

Head of Research Laboratory, MLU Halle-Wittenberg, Faculty of Medicine, University Hospital Halle

(Saale), Department of Orthopaedics, Trauma and Reconstructive Surgery, Laboratory for Experimental Orthopaedics & Sports Medicine, Ernst-Grube-Straße 40, 06120 Halle (Saale) download

13 February

2023

and

4 October

2023

Dr. Björn Zimmermann / Sports scientist and authorised signatory at h/p/cosmos

Dr Zimmermann has also created a PDF and a short video on the topic of energy transfer from electrically powered treadmills to the human body. Download PDF 20230213_Zimmermann_hpcosmos_Begruendung_Energietransfer_Laufband_Risikoklasse_II_EN_DE.pdf

image

Electrical energy into kinetic energy

image


20231004_Energy_transfer_treadmill_speed_incline.mp4

image


image

8 April 2011

Government of Upper Bavaria – Munich Trade Supervisory Office

[…] "The classification of medical devices is based on the principle that the higher the potential risk of negative effects on the human body, the higher the medical device is classified in the corresponding class. According to the recommendations of the EUROM VI and F+O working groups, based on the usual applications of medical devices, treadmill ergometers are assigned to class IIa in the list of classification examples in the "Ergometry" section.

Notwithstanding this classification, taking into account Section II No. 2.5 Annex IX

Classification Criteria 93/42/EEC and the potential risk to patients described by you within the meaning of Rule 10 - active diagnostic products specifically intended for monitoring vital physiological parameters and where the nature of the change could lead to an immediate risk (deliberately induced cardiac dysfunction) to a patient - endorses your classification as IIb." […]


The BfArM does not examine product groups, but always examines each product on a case-by-case basis, taking into account the respective intended purpose and other performance data. The applicability of Rule 9 and the fact that mechanical kinetic energy is transferred/exchanged on the treadmill has now also been confirmed by the BfArM, among others. Ergo, h/p/cosmos assumes that Rule 9 applies to all motorised treadmills because the basic laws of physics and biomechanical principles of walking and running on a treadmill are identical regardless of the manufacturer and model. Motorised treadmill = Class IIa or IIb.


Example photos of simple "standard treadmills", e.g. in physiotherapy:


image


We would like to quote from expert opinions (expertise in biomechanics):


Univ.-Prof. Dr. Hermann Schwameder

[…]

During the ground contact phase, both when running and walking, there is mechanical interaction between the locomotor and the moving treadmill belt. Accordingly, the kinetic energy of the electrically driven roller-treadmill belt system is transferred to the body of the locomotor, which then moves backwards. Against this background, an energy transfer (exchange of energy) takes place between the running belt, including the roller, and the person during the ground contact phase. Several studies show that this energy transfer during ground contact is not constant, but fluctuates due to mechanical conditions (braking and acceleration phases), which is particularly evident in the change in mechanical power during the ground contact phase and thus also indicates the aforementioned energy transfer between the treadmill and the locomotor (Savelberg et al., 1998; Sloot et al., 2014; Van Caekenberghe et al., 2012). The extent of these fluctuations depends, among other things, on body mass, treadmill mechanics and movement speed (Willwacher et al., 2021).

Both the theoretical principles and considerations as well as the empirical studies cited show that during locomotion on a treadmill, whether walking or running, there is a constant transfer of energy from the treadmill belt to the person and vice versa, i.e. an exchange of energy (energy transfer) between the treadmill and the locomotor.

[…] - End of quote –

Associate Professor Dr. phil. Rene Schwesig

[…]

  1. From a technical point of view, treadmills behave like conveyor belts (e.g. conveyor belts, luggage belts): they move objects from A to B. When a patient stands on the treadmill and it is started, the patient is moved backwards. The centre of mass (see above) is moved backwards. The patient must move to avoid falling. Similarly, the patient is moved upwards when the treadmill is driven into an incline by a lift motor. In both cases, kinetic energy is transferred from the treadmill to the patient.

  2. Although the energy transfer is clearly illustrated in point 1, standing on a treadmill is not the classic use. However, a hybrid form is a method of locomotion therapy in which one leg stands on the side footplates and the leg to be trained is placed on the moving treadmill belt (by the runner themselves or by the therapist) in order to move it passively backwards and thus mobilise it. The situation is similar when the patient is in a harness or lifting system (weight relief) and both legs are moved alternately through the running belt. This form is mostly used for patients with or without limited motor control of the legs (e.g. after a stroke or hemiplegia). Energy is thus transferred to the individual leg, which is partially or completely delivered by the treadmill.

  3. b. Speed changes during the gait cycle

Various studies have shown that treadmill speed is not constant during walking or running and is subject to variation (Nigg et al. 1995; Savelberg et al 1998; Willwacher et al. 2001). Motor power, control regulation, power supply, belt tension and lubrication, running style and runner weight play a decisive role here. When the runner's heel first makes contact with the treadmill ("heel strike"), the treadmill is initially slowed down and then accelerated again due to the motor control and the runner's push-off. Depending on the motor control, the speed may even exceed the set speed.

The runner thus has a significant effect on the belt speed, which Sloot and colleagues (2014) cite as the main reason for energy transfer (from the runner to the treadmill). Nevertheless, according to Savelberg and fellow researchers (1998), the continuous change in speed within a gait cycle causes energy to be transferred from the treadmill to the test subject. In their studies, Sloot and colleagues (2014) estimate an energy transfer of 0.11 joules per step at walking speeds. The running belt thus "pulls" the leg under the runner during the stance phase (Frishberg 1983).

See MDR Annex VIII Chapter II Implementing Rule 3.5. "If, taking into account the intended purpose specified by the manufacturer, several rules or several sub-rules within the same rule are applicable to one and the same product, the strictest rule/sub-rule shall apply, so that the product is classified in the highest class."


https://health.ec.europa.eu/system/files/2021-10/mdcg_2021-24_en_0.pdf

“In case several rules, or if, within the same classification rule, several sub-rules, apply to the same device based on the device intended purpose, the strictest rule and sub-rule resulting in higher classification will apply.”


A treadmill can therefore theoretically fall under several classification rules depending on its function, specification, intended purpose and/or integration into systems:

Rule 9 (active therapeutic products, e.g. stand-alone treadmill)

Rule 10 (active products for monitoring physiological parameters, e.g. ECG or stress test system = ECG with treadmill) Rule 11 (software)

If all components work together and provide medically relevant data, the entire system can be considered a medical device. See also MDR Article 22 "Systems and procedure packs".

The risk class is based on the highest risk component in the system – usually the ECG or the evaluation software.


Example: EMDN code: Z12050101: Stress analysis systems or GMDN code: 17877 Exercise stress testing system or GMDN code 17895: "Metabolic Carts, Stress Exercise / STRESS EXERCISE TROLLEY"

Definition: A system used to monitor cardiovascular function during physical exercise, typically including a treadmill or ergometer and ECG monitoring.


Regulatory guidance documents:


Medical Device Coordination Group Document MDCG 2021- 24

https://health.ec.europa.eu/latest-updates/mdcg-2021-24-guidance-classification-medical-devices-2021-10-04_en

1 Purpose of medical device classification

The classification of medical devices in use by the EU medical device legislation is a risk-based system taking into account the vulnerability of the human body and the potential risks associated with the devices. This approach uses a set of criteria that can be combined in various ways in order to determine classification, e.g. duration of contact with the body, degree of invasiveness, local vs. systemic effect, potential toxicity, the part of the body affected by the use of the device and if the device depends on a source of energy. The criteria can then be applied to a vast range of different medical devices and technologies. These are referred to as the ‘classification rules’ and are set out in Annex VIII of Regulation (EU) 2017/745 on medical devices (MDR). They correspond, to a large extent, to the classification rules established by the International Medical Device Regulators Forum (IMDRF) in the guidance document GHTF/SG1/N77:20121.

[…]


The MDCG guidance documents are considered binding by the authorities and notified bodies.


Examples of risk classes taken from the h/p/cosmos classification protocol document ID: 0684 rev. 1.4, supplemented by BfArM decision dated 28 February 2025 with reference number: 91.3.32-5634-K-0049/24


Risk Class

Medical Device Risk

Examples

CE marking

Notified Body


Class I


low risk

Hospital beds, walking aids, manual wheelchairs, parallel bars, expander training devices for rehabilitation


c


no

Class Is

sterile medical devices

sterile urine bags, wound dressing, etc.

c+4-digit no.

yes

Class Im

medical device with measuring function

scales, blood pressure monitor (non-active and non-invasive), etc.

c+4-digit no.

yes

Class Ir

reusable surgical instruments

surgical equipment sterilised and reused by hospitals, etc. ("r" stands for "reusable")

c+4-digit no.

yes


Class IIa


medium risk

treadmills with motor drive (with normal medium risk, based on BfArM decision 28.02.2025 reference 91.3.32-5634-K-0049/24), medical ergometers, electrocardiographs, electronic thermometers, electronic blood pressure measuring equipment, ultrasonic devices, external hearing aids, short-term corrective contact lenses, muscle stimulators, surgical gloves, dental filling materials and pins, medical device software intended to monitor physiological processes that are not considered to be vital, medical devices intended to be used to obtain readings of vital physiological signals in routine

check-ups, including monitoring at home


c+4-digit no.


yes


Class IIb


medium to high risk

treadmills with motor drive and high potential

risks, defibrillators, lung ventilators, patient monitors, X-ray, CT, etc.


c+4-digit no.


yes


Class III


high risk

stents, breast implants, implantable artificial joints, implantable heart pacemakers, heart valves, medical device software intended to perform diagnosis by means of image analysis for making treatment decisions in patients with acute

stroke, etc.


c+4-digit no.


yes


For medical devices in risk classes Is, Im, Ir, IIa, IIb and III, Article 52(6) of the MDR provides for a conformity assessment procedure that cannot be carried out solely by the manufacturer.


It is mandatory to involve a notified body within the meaning of Article 2 No. 42 of the MDR.


In such cases, the CE marking of medical devices, which is mandatory in any case, must be supplemented in accordance with Article 20(5) of the MDR by the four-digit identification number of the respective notified body with which the manufacturer has cooperated in the conformity assessment and which has issued the certificate establishing marketability. This CE marking,

including the identification number, must also be indicated in the manufacturer's declaration of conformity, in the instructions for use of the product and on the sales packaging. See Article 20(3) MDR.


It is also undisputed that the powerful motor drive of treadmills and the associated energy supply/energy exchange and the resulting strain on the patient pose a potential risk that is greater than that associated with a resting ECG application with an ECG of risk class IIa.


The potential risk exists in the use of mechanotherapy, gait therapy, physiotherapy and ergometry on the treadmill with ECG in the stress test function, e.g. in the BRUCE protocol.

Patients are often also restricted in their mobility (e.g. after a stroke) and/or in their ability to communicate, e.g. if a mouthpiece or face mask is also used for VO2max measurement/CPET.


See also relevant guidelines from or international publications: https://link.springer.com/content/pdf/10.1186/s12931-021-01895-6.pdf https://pmc.ncbi.nlm.nih.gov/articles/PMC6050434/pdf/Dtsch_Arztebl_Int-115_0409.pdf https://www.ahajournals.org/doi/pdf/10.1161/CIR.0b013e3181e52e69


Here an excerpt from the DGK ergometry guidelines from the German literature. Excerpt: https://leitlinien.dgk.org/files/2018_Manual_Stellenwert_Ergometrie.pdf

Or latest version: https://leitlinien.dgk.org/files/2025_manual_stellenwert_der_ergometrie.pdf


Requirements

Indication

As with all examination procedures, an ergometric examination must be based on a justified indication. This is based on the recommendations in Tables 6, 8, 9, 10, 11 and 12.

If, for example, it is already clinically and anamnestically foreseeable that ergometry will not be diagnostically useful for detecting ischaemia (e.g. patients with severe comorbidities or those in whom an increased risk of circulatory instability is to be expected), it should not be performed in the first place. In such cases, alternative (imaging) methods for diagnosing ischaemia should be used as a first resort [2, 3].


Contraindications to ergometry

Absolute and relative contraindications (CI) must be ruled out before the exercise test (Table 4). In the case of relative CI, ergometry should only be considered if the benefits and diagnostic yield are expected to significantly outweigh the potential risk.


Emergency equipment

For every exercise ECG, facilities for treating incidents, including resuscitation, must be available. A defibrillator is an essential part of the emergency equipment (must be checked regularly). One defibrillator per functional unit/practice is sufficient if all rooms in which ergometry is performed are immediately adjacent and therefore the access time to a defibrillator is equally short.

In summary, emergency equipment includes:

However, the intended user is authorised to allow the patient to operate the device in accordance with the instructions under constant supervision. This means that the operation of the device remains the responsibility of the intended user at all times, taking into account the physical and mental condition of the test person. The intended user must be within reach at all times (patient environment = 1.5 m radius).


Intended patient groups/target groups (excerpt as an example):

Adults and children > 1 year.

It is important to note that the decision to use the devices, with their potential risks and complications, for the diagnosis, rehabilitation or therapy of a specific patient is essentially the responsibility of the medical staff. The clinical user's assessment, on the other hand, must be based on the current state of medical science and the specific situation of the patient. The indications, target population and target user groups for treadmill tests and treatments must be decided by the physician and must be derived primarily from internationally recognised guidelines.


Intended duration of use (excerpt as an example):

Temporary: Normally intended for continuous use of less than 60 minutes. As specified by the physician.


Excerpt from the MDR:

EN: https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32017R0745&

Article 2 Definitions […]

(4)

‘active device’ means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices.

Software shall also be deemed to be an active device;

[…]

ANNEX VIII CLASSIFICATION RULES […] 6. ACTIVE PRODUCTS

    1. Rule 9

      All active therapeutic devices intended to administer or exchange energy are classified as class IIa unless their characteristics are such that they may administer energy to or exchange energy with the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy, in which case they are classified as class IIb.

      All active devices intended to control or monitor the performance of active therapeutic class IIb devices, or intended directly to influence the performance of such devices are classified as class IIb.

      All active devices intended to emit ionizing radiation for therapeutic purposes, including devices which control or monitor such devices, or which directly influence their performance, are classified as class IIb.

      All active devices that are intended for controlling, monitoring or directly influencing the performance of active implantable devices are classified as class III.



      […]

    2. Rule 10

      Active devices intended for diagnosis and monitoring are classified as class IIa:

      • if they are intended to supply energy which will be absorbed by the human body, except for devices intended to illuminate the patient's body, in the visible spectrum, in which case they are classified as class I;

      • if they are intended to image in vivo distribution of radiopharmaceuticals; or

      • if they are intended to allow direct diagnosis or monitoring of vital physiological processes, unless they are specifically intended for monitoring of vital physiological parameters and the nature of variations of those parameters is such that it could result in immediate danger to the patient, for instance variations in cardiac performance, respiration, activity of the central nervous system, or they are intended for diagnosis in clinical situations where the patient is in immediate danger, in which cases they are classified as class IIb.

      Active devices intended to emit ionizing radiation and intended for diagnostic or therapeutic radiology, including interventional radiology devices and devices which control or monitor such devices, or which directly influence their performance, are classified as class IIb.


      If a manufacturer incorrectly classifies a product in a lower risk class I, it has an unfair competitive advantage and is operating "under the radar" in regulatory terms.


      Here are some differences

      Requirements and significant costs

      Class I

      Class IIa

      Class IIb

      QM certification of the manufacturer by a notified

      body

      No

      Yes

      Yes

      Annual QM quality management system audits of the manufacturer by a notified body

      No

      Yes

      Yes

      Technical documentation creation

      Simple

      More complex

      More complex

      Clinical data CER Clinical Evaluation Report

      Simple

      More complex

      More complex

      Review of technical documentation by notified body prior to placing on the market

      No

      Yes

      yes

      Regular recurring review of technical documentation by notified body

      No

      Yes

      Yes

      Create and submit PSUR report / PMS report PSUR: Periodic Safety Update Report

      PMS: Post Market Surveillance

      PMS report "as required"

      PSUR

      "As needed", at least every 2 years

      PSUR

      "As required", at least annually

      Recipient of the PSUR or PMS report

      Authority (upon request)

      Authority (upon request) and notified body

      Authority (upon request) and Notified Body

      Person responsible for regulatory compliance. Person for Regulatory Compliance PRRC

      only under certain conditions

      with proven qualifications

      with proven qualifications

      These differences quickly add up to six-figure euro amounts annually (and even millions of euros for very large manufacturers and quantities) for internal employee costs and external costs, e.g. for independent testing laboratories with DAkkS accreditation, service providers, authorities and notified bodies (e.g. TÜV), and significantly reduce the time to market.

      Classification obligations according to the EN ISO 13485 standard

      1. Manufacturer's responsibility for classification

        ISO 13485 requires manufacturers to be familiar with and implement regulatory requirements, including the correct classification of the product in accordance with the MDR. Responsibility lies with the manufacturer, not the notified body.

      2. Documentation of classification

        The technical documentation must specify and justify the risk class of the product. ISO 13485 requires complete and traceable documentation, which also includes the classification rules applied (e.g. from Annex VIII MDR).

      3. Risk management in accordance with ISO 14971

        ISO 13485 refers to ISO 14971 for risk management. The risk class influences the depth and scope of the risk analysis. Incorrect classification can lead to inadequate risk assessment, which in turn violates ISO 13485.

      4. Conformity assessment and involvement of notified bodies

        Depending on the risk class, a notified body must be involved in the conformity assessment procedure. ISO 13485 requires that all regulatory requirements – including correct involvement – are met.

      5. Training and competence of employees

        The standard requires that all persons entrusted with regulatory tasks (e.g. classification, documentation, risk management) are appropriately trained and competent. Incorrect classification may indicate a lack of training and is a risk of non-compliance.

      6. Internal audits and corrective actions

        ISO 13485 requires regular audits. Incorrect classification must be identified as a deviation and corrected through corrective action. Repeated or deliberate violations can lead to the withdrawal of certification.


        ISO 13485 indirectly obliges manufacturers to classify risks correctly by requiring compliance with all regulatory requirements, complete documentation, effective risk management and competent employees. Deliberate or negligent downgrading of the risk class constitutes a violation not only of the MDR but also of ISO 13485 – with possible consequences such as audit deviations, loss of certification or market ban.


        What is the competitive advantage of incorrect classification (too low a risk class)?


        The incorrect classification of treadmills not only constitutes a violation of the EU Medical Devices Regulation (MDR) and the local implementation acts, such as the MPDG Medical Devices Implementation Act (German Act; similar implemented in many countries), but also unfair competition.

        Manufacturers of Class I products save a great deal of time and money each year and can bring medical devices to market much earlier without the need for an independent notified body.


        Obligations regarding the placing on the market, market surveillance and vigilance of medical devices (excerpt):

        Manufacturers (Art. 10 MDR)

        • Perform conformity assessment

        • Prepare technical documentation

        • Affix CE marking

        • Appoint a "Person Responsible for Regulatory Compliance"

        • Register with EUDAMED

        • Set up a post-market surveillance system

        • Regularly update safety reports

        • Analyse and report incidents and trends

        • Initiate corrective measures

          Importer (Art. 13 MDR) or distributor (Art. 14 MDR)

        • Check whether the product is CE marked

        • Check manufacturer information

        • Affix name and address to the product or packaging

        • Register the product in EUDAMED

        • Ensure traceability

        • Check conformity before provision (e.g. CE marking, instructions for use, declaration of conformity)

        • Ensure storage and transport

        • Has the importer covered up the manufacturer's markings with their additional (!) markings?

        • Has the manufacturer assigned the UDI?

        • Does the product appear to comply with legal requirements?

        • Forward complaints and incidents and cooperate with authorities

        • Report incidents to manufacturers and authorities

        • Maintain a register of non-compliant products

        • Cooperate with recalls and safety measures

        • Keep documentation available for traceability


Sources:

https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32017R0745#anx_VIII https://www.medical-device-regulation.eu/mdr-article-14-general-obligations-of-distributors/ https://health.ec.europa.eu/medical-devices-sector/directives/market-surveillance-and-vigilance_en


Motorised treadmills are also machines:

Another important fact for manufacturers and distributors to note is that motorised treadmills are subject to the EU Machinery Regulation (EU) 2023/1230 (formerly Machinery Directive 2006/42/EC) in addition to the MDR for medical devices.

This also requires a declaration of conformity for machines, the preparation of technical documentation and the appointment of a CE representative with the designation of a person responsible for technical documentation and conformity.


The EU Machinery Regulation also clearly regulates essential responsibilities within the European Union. Examples:

CHAPTER II

OBLIGATIONS OF ECONOMIC OPERATORS

Article 10

Obligations of manufacturers of machinery and related products

[…]

  1. When placing machinery or a related product on the market or putting it into service, manufacturers shall ensure that it has been designed and constructed in accordance with the essential health and safety requirements set out in Annex III.

  2. Before placing machinery or a related product on the market or putting it into service, manufacturers shall draw up the technical documentation set out in Annex IV, Part A and carry out the relevant conformity assessment procedure referred to in Article 25 or have it carried out.

Where compliance of machinery or a related product with the essential health and safety requirements laid down in Annex III has been demonstrated by that conformity assessment procedure, manufacturers shall draw up the EU declaration of conformity in accordance with Article 21 and affix the CE marking in accordance with Article 24.

[…]


Article 15

Obligations of distributors of machinery and related products

  1. When making machinery or a related product available on the market, distributors shall act with due care in relation to the requirements of this Regulation.

  2. Before making machinery or a related product available on the market, distributors shall verify that:

(a)

the machinery or related product bears the CE marking;

(b)

the machinery or related product is accompanied by the EU declaration of conformity referred to in Article 10(8);

[…]


https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32023R1230

Applicability of the Machinery Directive according to MDR/IVDR

"Devices that are also machinery within the meaning of point (a) of the second paragraph of Article 2 of Directive 2006/42/EC of the European Parliament and of the Council (35) shall, where a hazard relevant under that Directive exists, also meet the essential health and safety requirements set out in Annex I to that Directive to the extent to which those requirements are more specific than the general safety and performance requirements set out in Chapter II of Annex I to this Regulation."

Source: Art. 1(12) MDR; Art. 1(6) IVDR


Interesting article on this topic: https://blog.johner-institute.com/systems-engineering/machinery-directive-medical-devices/


The EU Machinery Regulation 2023/1230 replaces the EU Machinery Directive 2006/42/EC, and the articles will come into force on 13 July 2023, with all articles coming into full force on 14 January 2027.

Just as cybersecurity has played an enormously important role in medical devices since the MDR, the new Machinery Regulation also addresses cybersecurity, and manufacturers will have to protect their machines against cyber threats in future. This mainly concerns cyber attacks, but also hardware and software failures, disrupted wireless connections and errors in AI-based autonomous operation functions.


The legislator transposes the EU directives and regulations on CE marking into national law. Companies are obliged to ensure conformity in all CE processes.

The role of the CE representative is central to this. In Germany, this is regulated, among other things, by the Act on the Provision of Products on the Market* (Product Safety Act – ProdSG), which refers to the Machinery Directive/Machinery Regulation. https://www.gesetze-im-internet.de/prodsg_2021/


Summary:


  1. Motor-driven treadmills as medical devices must always be classified as Class IIa or Class IIb in accordance with (EU) 2017/745 MDR Rule 9 and, if applicable in accordance with Sub-rule 10. Motorized treadmills are always also machines.

  2. Manufacturers of treadmills must be certified by accredited/certified independent notified bodies in accordance with the MDR, and their quality management systems, technical documentation, PSUR reports with PMS market surveillance and clinical data (CER file) must be monitored regularly.

    Comparison: Just as a person cannot issue their own driving license in the form of a "self-declaration" without having successfully passed a test (even if they think they can drive a car) and/or a technically skilled person cannot issue a TÜV sticker in a "self-declaration" (even if they think their car is safe), manufacturers and their QM systems, including risk management, clinical data and TecDoc for all Class IIa and/or IIb medical devices, must be certified and monitored annually by independent and accredited bodies.

  3. According to Article 5(1) of the MDR, a medical device may only be placed on the market or put into service if it complies with the requirements of the MDR. The same applies to requirements for machinery under (EU) 2023/1230.

  4. Incorrect classification of a medical device and/or non-compliance with the EU Machinery Directive means that the product is not marketable and therefore may not be distributed or placed on the European market. Advertising and selling a non- marketable product is not permitted.

  5. Violation of the provisions of Articles 13 and 14 MDR and also of the provisions of the EU Machinery Regulation 2023/1230 also constitutes a breach of competition law within the meaning of Section 3a UWG (German Unfair Competition Act).

  6. Economic operators such as distributors and importers also have a duty to cooperate in checking the manufacturer's documents and are jointly responsible for ensuring that the relevant EU declarations of conformity accompany the products.

  7. Anyone who deliberately or negligently places a machine or safety component on the market without an EC declaration of conformity or CE marking is acting unlawfully (Section 8 No. 9 ProdSV). The offence can be punished with a fine of up to one hundred thousand euros (cf. Section 28 (2) Product Safety Act (ProdSG)). Link

  8. The state authorities, regional councils, trade supervisory offices and health authorities of the federal states are responsible for operational market surveillance, can carry out random checks, product tests and inspections at manufacturers, distributors and importers, and can assess complaints and incidents.


    Author: Franz Harrer

    h/p/cosmos sports & medical gmbh

    Co-founder, shareholder and expert in technology, risk management and regulatory affairs since 1988 Appendix: Web findings on court ruling on medical device with incorrect risk class or without EU conformity.

    Web findings:

    There have been initial court rulings in which a medical device was incorrectly classified, leading to legal consequences. Here is a particularly relevant case:

    1. Skin check app – incorrect classification as Class I

      • Court: Hanseatic Higher Regional Court of Hamburg

      • File number: 3 W 30/23

      • Date: 22 September 2023

      • Facts of the case:

        A dermatological software app for examining skin changes was classified as risk class I by the manufacturer. However, the court ruled that the app provides medical information for diagnosis and therefore belongs in at least class IIa according to Rule 11 MDR.

      • Consequences:

        • The app would have required a notified body for CE certification.

        • The incorrect classification led to a competition dispute with a competitor.

        • The court emphasised the importance of the intended purpose and actual functionality of the product.

          Conclusion:

          Incorrect risk classification can lead to:

      • legal disputes,

      • sales bans,

      • financial losses and

      • reputational damage.

      The intended purpose, technical function and actual use are crucial for correct classification. Source / References:

      https://www.landesrecht-hamburg.de/bsha/document/NJRE001563508


      image


    2. Teeth whitening products as Class Ila medical devices. Court: Hanover Regional Court, judgement of 18 July 2001 - 22 O 1075/01 Competition law status of the European Authorised Representative (EAR) […]

      Facts of the case:

      The plaintiff demands that the defendant, as EAR for two tooth whitening products from a Californian importer, refrain from marketing these products without the certification procedures required for classification as a Class Ila medical device; the plaintiff itself markets products similar in application and effect as Class IIa medical devices with the associated certification procedures. […]

      Result: The EAR of an American manufacturer may not permit the anti-competitive distribution of a medical device due to incorrect classification; rather, it may be held liable under competition law for failing to refrain from placing the medical device on the market without the required certification.

      […]

      The action is well founded; the plaintiff may demand that the defendant refrain from marketing the disputed products without the certification required for a Class Ila medical device.

      The right to injunctive relief arises from Section 1 of the Unfair Competition Act (UWG). Source: Medical Devices Journal 9th year · Issue 2 · 2002


      image

    3. State Office for Health and Occupational Safety North Rhine-Westphalia

Excerpt from the article: "If the supervisory authority determines that a machine is being operated without the required declaration of conformity and CE marking, the employer may face a shutdown order until the machine is brought into compliance with the law.

If an accident occurs on such a machine, criminal investigations may be launched against both the distributor and the employer."

Source: https://www.komnet.nrw.de/_sitetools/dialog/5364